Fluke 9100 User Handbook Manual page 151

Universal calibration system
Hide thumbs Also See for 9100:
Table of Contents

Advertisement

8.3.4.4
Display Checks
Display checks are initiated by pressing the DISPLAY key on the 'Select test' menu
screen:
Selftest.
Select test using softkeys
TODAYS DATE
E X I T
Selecting 'DISPLAY' transfers to the 'Display test.' screen. This invites a user to use the
front panel'tab' ( ) key to move the dark band down the screen, up to the top, and down
again, thus testing all the display elements on the screen:
Display test.
Use the tab key to advance the band
and the enter key to exit the test.
The screen text will shift automatically so as not to obscure the band as it jumps back
to the top of the screen. If there are elements of the screen which do not show light in
the light background or dark in the dark band as it is advanced down the screen, then this
implies a failure. Rectification will require access to the internal circuitry, so no further
user action is recommended, except to report the result to your Fluke Service Center.
The '
Enter
' Key ( )
returns to the Interface '
Section 8: Model 9100 — Routine Maintenance and Test
D I S P L A Y
M E M O R Y
K E Y B R D
D I S P L A Y
M E M O R Y
C A R D
T R A C K E R
P R I N T E R
TIME
Select test
' menu screen.
Final Width = 215mm
8-19