Fluke 9100 User Handbook Manual page 145

Universal calibration system
Hide thumbs Also See for 9100:
Table of Contents

Advertisement

8.3.2.2
Selftest Runs to Completion
If the selftest is not aborted, it will run to completion, and if the test is successful with
no failures, the following screen will appear for either Fast and Full test:
Selftest.
Test completed with no failures.
TODAYS DATE
E X I T
P R I N T
If failures were encountered during the test, the 9100 will display the following screen:
Selftest.
Selected test has FAILED.
Number of failures: XX
Use the softkeys to view the results.
TODAYS DATE
E X I T
P R I N T
EXIT
returns to the '
Select selftest
prints out the results of the test, whether Full or Fast.
PRINT
Refer to Sub-section 8.4.
If failures were encountered,
EXIT
and
failures is shown on the screen, and an extra selection will be available:
sets up a special screen for detailing the parameters of the failures
VIEW FAILS
encountered (described later in Paras 8.3.2.3).
Section 8: Model 9100 — Routine Maintenance and Test
TIME
TIME
VIEW
FAILS
' menu screen.
PRINT
remain available. Also, the number of
Continued Overleaf
Final Width = 215mm
8-13