8.3.1.1
Entry to Test Mode
The Mode key sets up a special menu display, offering selection from five primary
modes. This menu can be exited only by pressing one of the five screen keys.
Mode Selection
Hz
Select required mode using softkeys.
Aux
Mode
TODAYS DATE
PROC
TEST
This key enters Test mode, displaying the following screen:
Selftest
Select required test using softkeys.
TODAYS DATE
F A S T
FAST
runs a
FAST
selftest.
FULL
runs a
FULL
selftest.
INTERFACE
allows checks of the display and display memory, the front panel
keyboard, the (Procedure mode) memory card slots, the mouse, and the
printer interfaces.
Section 8: Model 9100 — Routine Maintenance and Test
(Contd.)
TIME
MANUAL CONFIG
CALIB
TEST
TIME
INTER-
F U L L
FACE
Continued Overleaf