Fluke 9100 User Handbook Manual page 78

Universal calibration system
Hide thumbs Also See for 9100:
Table of Contents

Advertisement

Final Width = 215mm
6.C.2
Event Status Enable
This event status data structure conforms to the IEEE 488.2 standard requirements for this structure.
ESE enables the standard defined event bits which will gener-
ate a summary message in the status byte. Refer to Section 6;
Subsection 6.5.
Nrf is a Decimal Numeric Data Element representing an integer
decimal value equivalent to the Hex value required to
enable the appropriate bits in this 8-bit register. The
detailed definition is contained in the IEEE 488.2 standard
document. Note that numbers will be rounded to an
integer.
6.C.3
Recall Event Status Enable
This event status data structure conforms to the IEEE 488.2 standard requirements for this structure.
ESE?
recalls the enable mask for the standard defined events. Refer
to Section 6; Subsection 6.5.
Response Decode:
The value returned, when converted to base 2 (binary), identifies
the enabled bits which will generate a summary message in the
service request byte, for this data structure. The detailed
definition is contained in the IEEE 488.2 document.
6-C2
Appendix C to Section 6: 9100 System Operation — IEEE 488.2 Common Commands and Queries
ESE
Nrf
phs
Execution Errors:
None.
Power On and Reset Conditions
Not applicable.
ESE?
Execution Errors:
None
Power On and Reset Conditions
The Power On condition depends on the condition stored by the
common PSC command - if 0 then it is not cleared; if 1 then the
register is cleared. Reset has no effect.

Hide quick links:

Advertisement

Table of Contents
loading