6.6.3.4
OUTP:ISEL(?)
<CPD>{HIGHi|HI50|HI10|LOWi}
Purpose
The I+ Output in Current function can be passed out via the unguarded front panel I+ terminal, or on a guarded line (pin 8) through
the D-type socket ('Signal Output' J109) beneath the terminals. Pin 7 of J109 is connected to internal Guard shields for this purpose,
and is available for external guard connection. The return is always via the front panel main I- terminal.
The front panel I+ terminal can carry all DC currents and AC currents up to the maximum -20A to +20A output (refer to
Volume 1, Section 4, paras 4.5.5.3 for time limitations). Pin 8 of J109 is limited to -1A to +1A output.
OUTP:ISEL HIGH
switches current outputs via the front panel I+ terminal,
(Option 200) switches current outputs via the front panel I+ terminal, and provides equivalent
OUTP:ISEL HI50
currents of (16.0A to 1000A) into the 50-turn current coil.
OUTP:ISEL HI10
(Option 200) switches current outputs via the front panel I+ terminal, and provides equivalent
currents of (3.2A to 200A) into the 10-turn current coil.
switches current outputs via J109 pin 8.
OUTP:ISEL LOW
Response to Query Version
The 9100 will return HIGH/HI50/HI10/LOW to match the active programming parameter.
Section 6: 9100 System Operation — SCPI Language
Final Width = 215mm
6-29