10.4.2.2
Sequencing Calibrations
The table below indicates the order in which the various Model 9100 functions should be
calibrated. Although it is not essential to calibrate all the functions indicated below at any
one time, functions higher in the list should be calibrated before those lower in the list.
If Resistance and Capacitance are being calibrated separately, the Resistance function
must be calibrated before the Capacitance function. Similarly if the Resistance
function's 'super' UUTi, 'high' UUTi and 'low' UUTi modes are being calibrated, some of
the 'super' UUTi hardware, then 'high' UUTi hardware configurations must be calibrated
before associated 'low' UUTi hardware configurations.
Sequence
Function
1
DC Voltage (Zero to 320V)
2
AC Voltage (Zero to 320V)
3
Resistance (Super UUTi)
4
Resistance (High UUTi)
5
Resistance (Low UUTi)
6
DC Voltage (320V to 1050V)
7
AC Voltage (320V to 1050V)
8
Capacitance
9
DC Current (Zero to 320mA)
10
AC Current (Zero to 320mA)
11
DC Current (0.32A to 20A)
12
AC Current (0.32A to 20A)
Table 10.4.2.1: Recommended Sequence of Calibrations
Section 10: Calibrating the Model 9100: Summary of Calibration Process
Reference
Section 10.4.3
Section 10.4.4
Section 10.4.7
Section 10.4.7
Section 10.4.7
Section 10.4.3
Section 10.4.4
Section 10.4.8
Section 10.4.5
Section 10.4.6
Section 10.4.5
Section 10.4.6
10.4-5