6.C.12
Reset
RST
will reset the instrument to a defined condition, stated for each
applicable command with the command's description, and listed
in Appendix D to Section 6.
The reset condition is not dependent on past-use history of the
instrument except as noted below:
RST does not affect the following:
the selected address of the instrument;
calibration data that affect specifications;
SRQ mask conditions;
the state of the IEEE 488.1 interface.
6.C.13
Service Request Enable
This Status Byte data structure conforms to the IEEE 488.2 standard requirements for this structure.
SRE enables the standard and user-defined summary bits in the service request byte, which will generate a service request. Refer
to Section 6, Subsection 6.5.
Nrf
is a Decimal Numeric Data Element representing an
integer decimal value equivalent to the Hex value required
to enable the appropriate bits in this 8-bit register. The
detail definition is contained in the IEEE 488.2 document.
Note that numbers will be rounded to an integer.
Appendix C to Section 6: 9100 System Operation — IEEE 488.2 Common Commands and Queries
*
RST
Execution Errors:
None.
Power On and Reset Conditions
Not applicable.
*
SRE
phs
Execution Errors:
None.
Power On and Reset Conditions
Not applicable.
Nrf