Fluke 9100 User Handbook Manual page 125

Universal calibration system
Hide thumbs Also See for 9100:
Table of Contents

Advertisement

7.19.7 AC Current Accuracy (sine)
Output Current
0 to 320.0mA
0.320 to 3.200A
3.200 to 10.500A
10.50 to 20.00A
3.20 to 32.00A, 10t coil
32.0 to 200.0A, 10t coil
16.0 to 160.0A, 50t coil
160.1 to 1000A, 50t coil
Note that above specs are subject to maximums as specified in the Amp.Hz profile.
Performance above 1kHz not specified to user; nominally Gain is as 9100 and Floors as above.
All other specifications as 9100 except that the 0.2% adder for the Option 200 coils has been included in the table above.
Also directly available as a voltage harmonic output. When used as a harmonic with n>1, add 0.04% to percentage of output specifications.
7.19.8 Other Current Waveshapes
All subject to maximum frequencies as per the main 9100 specification tables.
Ranges available start at the 45.25mA to 452.5mA pk range.
This is allowed to function down to zero.
Performance as in main 9100 specifications.
Frequency
Accuracy
Band
(% Output + Floor)
10 - 1k
0.08 + 32 A
10 - 1k
0.08 + 480 A
10 - 100
0.11 + 3mA
100 - 1k
0.15 + 3mA
10 - 100
0.11 + 6.9mA
100 - 1k
0.15 + 6.9mA
10 - 100
0.40 + 9.7mA
100 - 440
0.98 + 27mA
10 - 100
0.40 + 90mA
100 - 440
0.87 + 250mA
10 - 100
0.42 + 48mA
10 - 100
0.42 + 450mA
Section 7: Model 9100: Specifications
Final Width = 215mm
7-29