10.4.4.4
Calibration Setup
1. Connections
Ensure that the 9100 is connected to the Standards DMM as shown in Fig. 10.4.4.1, and that both instruments
are powered on and warmed up.
2. 9100
a. Ensure that the 9100 is in STANDARD CAL mode and then select the AC Voltage function by pressing
the 'V' key on the right of the front panel, followed by pressing the ' V' screen key adjacent to the display.
b. Ensure that Sinusoidal waveshape is selected, and that the Phase facility is deselected (
lettering on light background).
Table 10.4.4.1: AC Voltage (Zero to 320V) Hardware Configurations and Calibration Points
Output span
Output
covered by
to select
hardware
hardware
configuration
config.
Default
000.000mV
100mV
300.000mV
to
300.000mV
320.000mV
300.000mV
0.32001V
1V
3.00000V
to
3.00000V
3.20000V
3.00000V
3.2001V
10V
30.0000V
to
30.0000V
32.0000V
30.0000V
32.001V
100V
100.000V
to
100.000V
105.000V
100.000V
105.001V
150V
300.000V
to
300.000V
320.000V
300.000V
300.000V
Table 10.4.4.2: AC Voltage (320V to 1050V) Hardware Configurations and Calibration Points
Output span
Output
covered by
to select
hardware
hardware
configuration
config.
Default
320.01V
500V
750.00V
to
750.00V
1050.00V
750.00V
330.00V
10.4-16
Section 10: Calibrating the Model 9100: AC Voltage Function
Calibration Targets
Amplitude
Minimum
Maximum
Default
180.000mV
320.000mV
1.0000kHz
180.000mV
320.000mV
30.000kHz
180.000mV
320.000mV
100.00kHz
1.80000V
3.20000V
1.0000kHz
1.80000V
3.20000V
30.000kHz
1.80000V
3.20000V
100.00kHz
18.0000V
32.0000V
1.0000kHz
18.0000V
32.0000V
30.000kHz
18.0000V
32.0000V
100.00kHz
70.000V
105.000V
1.0000kHz
70.000V
105.000V
30.000kHz
70.000V
105.000V
100.00kHz
180.000V
320.000V
40.00Hz
180.000V
320.000V
200.00Hz
180.000V
320.000V
10.000kHz
180.000V
320.000V
30.000kHz
Calibration Targets
Amplitude
Minimum
Maximum
Default
700.00V
1050.00V
40.00Hz
700.00V
1050.00V
200.00Hz
700.00V
1050.00V
10.000kHz
320.10V
350.00V
30.000kHz
label dark
Factor
Number
Frequency
Minimum
Maximum
150.00Hz
5.000kHz
1
15.000kHz
45.00kHz
2
65.00kHz
100.00kHz
3
150.00Hz
5.000kHz
1
15.000kHz
45.00kHz
2
65.00kHz
100.00kHz
3
150.00Hz
5.000kHz
1
15.000kHz
45.00kHz
2
65.00kHz
100.00kHz
3
150.00Hz
5.000kHz
1
15.000kHz
45.00kHz
2
65.00kHz
100.00kHz
3
40.00Hz
75.00Hz
1
150.00Hz
250.00Hz
2
3.0000kHz
20.000kHz
3
25.000kHz
30.000kHz
4
Factor
Number
Frequency
Minimum
Maximum
40.00Hz
75.00Hz
1
150.00Hz
250.00Hz
2
3.0000kHz
20.000kHz
3
25.000kHz
30.000kHz
4