6.C.8
Power On Status Clear
This common command conforms to the IEEE 488.2 standard requirements.
PSC
sets the flag controlling the clearing of defined registers at
Power On.
Nrf is a decimal numeric value which, when rounded to an
integer value of zero, sets the power on clear flag false. This
allows the instrument to assert SRQ at power on, providing that
the PON bit in the ESR is enabled at the time of power-down, by
the corresponding bit in its Enable register (ESE).
When the value rounds to an integer value other than zero it sets
the power on clear flag true, which clears the standard event
status enable and service request enable registers so that the
instrument will not assert an SRQ on power up.
Appendix C to Section 6: 9100 System Operation — IEEE 488.2 Common Commands and Queries
PSC
Nrf
phs
Examples:
PSC 0 or PSC 0.173 sets the instrument to assert an SRQ at
Power On.
PSC 1 or PSC 0.773 sets the instrument to not assert an SRQ
on Power On.
Execution Errors:
None.
Power On and Reset Conditions
Not applicable.