Fluke 9100 User Handbook Manual page 233

Universal calibration system
Hide thumbs Also See for 9100:
Table of Contents

Advertisement

DANGER
HIGH VOLTAGE
sHi
sLo
15Vpk
max
!
!
Hi
Lo
1500Vpk
max
I+
I-
15Vpk
max
Coaxial Cable
e.g. 1513
Standards DMM
Guard
Guard
I+
I-
!
Hi
Lo
Interconnections for calibration points
Fig. 10.4.5.3
below 1.9A
Shunt should
typically produce
between 0.1V and
0.2V at cal point
I+
Shunt
e.g. 4953
Hi
Reset
Interconnections for calibration points
Fig. 10.4.5.4
Section 10: Calibrating the Model 9100: DC Current Function
HIGH VOLTAGE
DANGER
sHi
sLo
15Vpk
max
!
!
Hi
Lo
1500Vpk
max
I+
I-
15Vpk
max
I-
Case
Lo
Standards DMM
Guard
Guard
I+
I-
!
Hi
Lo
Reset
above 1.9A
Final Width = 215mm
10.4-25