Final Width = 215mm
Table 10.4.6.3: Auxiliary AC Current Calibration Points
Output span
covered by
hardware
configuration
Default
0.000mA to 320.0mA
300.000mA
(one range)
300.000mA
300.000mA
300.000mA
0.3201A to 3.200A
3.00000A
3.00000A
3.00000A
3.00000A
3.21 to 20.00A
10.0000A
10.0000A
10.0000A
10.0000A
10.4-34
Section 10: Calibrating the Model 9100: AC Current Function
Calibration Targets
Amplitude
Minimum
Maximum
Default
180.000mA
320.000A
40Hz
180.000mA
320.000A
150Hz
180.000mA
320.000A
1000Hz
180.000mA
320.000A
3000Hz
1.80000A
3.20000A
40Hz
1.80000A
3.20000A
150Hz
1.80000A
3.20000A
1000Hz
1.80000A
3.20000A
3000Hz
7.00000A
10.5000A
40Hz
7.00000A
10.5000A
150Hz
7.00000A
10.5000A
1000Hz
7.00000A
10.5000A
3000Hz
Frequency
Minimum
Maximum
35Hz
45Hz
140Hz
160Hz
900Hz
1100Hz
2700Hz
3000Hz
35Hz
45Hz
140Hz
160Hz
900Hz
1100Hz
2700Hz
3000Hz
35Hz
45Hz
140Hz
160Hz
900Hz
1100Hz
2700Hz
3000Hz