Fluke 9100 User Handbook Manual page 159

Universal calibration system
Hide thumbs Also See for 9100:
Table of Contents

Advertisement

Serial No: XXXXXX
S/W Issue: X.XX
TEST PT
MEAS VALUE
MAX LIMIT
AØ4.ØØ1
-Ø.ØØØØ44
+Ø.ØØ15ØØ
AØ1.ØØ1
+6.55334Ø
+6.6Ø28ØØ
PØ5.ØØ1
+5.115ØØ5
+5.725ØØØ
Serial No: XXXXXX
S/W Issue: X.XX
TEST PT
MEAS VALUE
MAX LIMIT
AØ1.ØØ1
+6.553292
+6.564493
AØ1.ØØ2
+6.553137
+6.5662ØØ
AØ1.ØØ3
-6.553888
-6.537325
8.4.2
Printing Setup
The results of Full and Fast selftest operations can be printed directly via the rear panel
parallel port J103. A suitable printer must be connected and switched on-line; then the
9100 internal program will generate the required results.
8.4.2.1
Printer Type
The printer to be used should be capable of printing 120 characters per line, and must
be able to print the Code Page 437 character set. Most printers compatible with Epson
FX, Canon Bubble-Jet or Hewlett-Packard Desk-Jet are suitable.
8.4.2.2
Data Formatting
The required printout style, format and data is pre-determined depending on whether
Full or Fast test results are being printed, and on the type of printer to be used. It is
necessary to enable a particular printer type via Configuration mode, only if the format
for that printer is required.
Note: If the printer is disabled (Config mode — Volume 1, Section 3, paras 3.3.2.6),
the printer will not be set up when starting to print for the first time.
8.4.3
Results Printout
Apart from the heading, both Full and Fast test results have the same printed layout. The
results of tests on all test pathways are collected together in a table. Typical samples of
tables are given below:
Date: XX/YY/ZZZZ
Time: 12:41
MIN LIMIT
NOMINAL
RESULT A
-Ø.ØØ15ØØ
Ø.ØØØØØØ
Ø.ØØØØ41
+6.5Ø44ØØ
+6.5536ØØ
+6.553363
+4.675ØØØ
+5.2ØØØØØ
+5.114981
Date: XX/YY/ZZZZ
Time: 12:41
MIN LIMIT
NOMINAL
RESULT A
+6.552743
+6.55367Ø
+6.553292
+6.541ØØØ
+6.5536ØØ
+6.553173
-6.569875
-6.5536ØØ
-6.553864
gives the achieved percentage of full tolerance for that test.
ERROR %
In this column a failure is shown by the figure '1' against the relevant test.
FAILURES
Section 8: Model 9100 — Routine Maintenance and Test
Type: FAST
RESULT B
ERROR %
FAILURES
-Ø.ØØØØ48
-3
+6.553316
Ø
+5.115Ø29
-16
Type: FULL
RESULT B
ERROR %
FAILURES
+6.553292
-3
+6.553181
-4
-6.553912
-2
Final Width = 215mm
8-27