There are two methods of operating the STP pump: MANUAL and
REMOTE.
Select one which fits your vacuum equipment.
7.4

Manual Operation

Slide the "MANUAL/REMOTE" changeover switch on the STP control unit front panel to the
position opposite to "ON". At this time, the "REMOTE" LED is off.
7.4.1

Powering ON

1)
Switch "ON" the breaker on the rear panel.
To prevent incorrect operation, a metal fitting is attached to the breaker. Loosen the
screw, lift the metal fitting, and secure it.
The magnet bearing functions and the rotor levitates (POWER ON state).
The "POWER" LED illuminates.
2)
The STP control unit executes the self test. If no error is found, the LCD displays
"LEVITATION".
7.4.2

Starting the STP Pump

1)
Press the "START" switch on the front panel to start and accelerate the STP pump. The
"ACCEL." LED illuminates. The LCD displays the current rotational speed in "ACCEL:
***00rpm" in units of 500 rpm (ACCELERATION state).
2)
When the STP pump attains the rated speed of rotations, the "ACCEL." LED
extinguishes, and the "NORMAL" LED illuminates. Also, the LCD displays "NORMAL:
***00rpm" (NORMAL OPERATION state).
◇ The STP pump can be manually started only when the LCD displays "LEVITATION"
or "BRAKE: ***00rpm".
In the case of another message displayed, press the "SELECT" switch one or more
times until "LEVITATION" or "BRAKE: ***00rpm" is displayed.
◇ Although the STP pump sounds for several seconds while it is accelerating, this is
not an indication of abnormality.
7.4.3

Stopping the STP Pump

1)
Press the "STOP" switch on the front panel to stop the STP pump.
The "NORMAL" LED or "ACCEL." LED extinguishes, and the "BRAKE" LED
illuminates (BRAKE state).
The "BRAKE" LED does not illuminate when the number of rotations is less than 500
rpm.
2)
The "BRAKE" LED extinguishes when the number of rotations goes below 500 rpm.
◇ The STP pump can be stopped with any message displayed on the LCD.
STP-L301/L451 Series Instruction Manual
7-3