Remote Output Signals; Table 8.7 Con 4 Connector (Remote Output Signal) - Edwards STP-L301 Series Instruction Manual

Turbomolecular pumps
Hide thumbs Also See for STP-L301 Series:
Table of Contents

Advertisement

8.3.2

Remote Output Signals

Use output signals according to Table 8.7 and Figure 8.4.
This output signal operates in both manual operation and remote operation.
No.4 of the DSW-B DIP switch operates in "ON/OFF" both states.

Table 8.7 CON 4 Connector (Remote output signal)

Pin
(5)
(17)
REMOTE
N.O OUT
(6)
(18)
ROTATION
N.O OUT
(7)
(19)
EMERGENCY
OPERATION
N.O OUT
(8)
(20)
FAILURE
N.O OUT
(9)
(21)
OVER
TEMPERATUR
N.O OUT
(11)
(23)
ACCELERATION
N.O OUT
(11)
(24)
NORMAL
OPERATION
N.O OUT
(10)
(22)
(12)
STP-L301/L451 Series Instruction Manual
Pins for outputting the STP pump REMOTE SELECTION state
signal.
When the MANUAL/REMOTE changeover switch on the STP front
panel is set to ON (remote side), these pins are closed (REMOTE
operation).
However, during power failure backup operation, this signal is not
output.
Pin for outputting the rotational state signal.
These pins are closed when the STP pump is in rotation.
However, during power failure backup operation, this signal is not
output.
(It corresponds to the pin for outputting the "POWER ON" signal of
the STP-200/300/400 series)
Pins for outputting the EMERGENCY OPERATION state signal.
These pins are closed when failures are detected as follows.
Abnormal state of magnetic bearing
Excessive vibration
However, during power failure backup operation, this signal is not
output.
Pins for outputting the FAILURE state signal.
These pins are closed when failures are detected as follows.
Motor driver overload
Overheating inside the STP pump
Overheating inside the STP control unit
Overspeed
However, during power failure backup operation, this signal is not
output.
Pins for outputting the OVER TEMPERATURE state signal.
These pins are closed when overheating inside the STP pump is
detected.
However, during power failure backup operation, this signal is not
output.
Pins for outputting the ACCELERATION state signal.
These pins are closed when the STP pump is in acceleration.
Pins for outputting the NORMAL OPERATION state signal.
These pins are closed when the STP pump is in rated operation.
Not used.
(It corresponds to the pin for outputting the "BATTERY
OPERATION " signal of the STP-200/300/400 series)
Not used. It is connected to the frame ground.
8-9
Description