Air Cooling Method - Edwards STP-L301 Series Instruction Manual

Turbomolecular pumps
Hide thumbs Also See for STP-L301 Series:
Table of Contents

Advertisement

5.2.2

Air Cooling Method

When water cooling is not available, use an air cooling unit (optional accessory).
Attach the air cooling unit using four screws at screw holes for legs (8-M8
(For the positions of the screw holes for legs, see "External Appearance of the STP Pump")
◇ Since the air cooling unit is attached using four screw holes for legs, the base for
securing the STP pump cannot be installed using screw holes for legs (see Section
4.3.3, "How to Secure the STP Pump"). When using the air cooling unit, design your
vacuum equipment so that the STP pump installing flange at the vacuum equipment
side can withstand the torque generated by the abnormal STP pump.
◇ Check the rated voltage of the air cooling unit before use.
◇ Procure a breaker or fuses for the air cooling unit (remote terminal block TB1
incorporates fuses).
◇ When connecting the power for the air cooling unit to the remote terminal block (I/O
TB1), attach a metallic cover or a similar part to the terminal block to ensure the
safety.
◇ DO NOT apply excessive force to the cable for the air cooling unit.
◇ The air cooling unit becomes vibration source.
Follow the steps below to run the STP pump associated with the air cooling unit (read
through Section 8, "Remote Input/Output Signal Terminal Blocks", then operate them.)
1)
Connect the power cable of the air cooling unit between "I/O TB1 FAN OUT Terminals
(3)-(4)".
2)
Input the power supply for the air cooling unit between "I/O TB1 AC. POWER IN
Terminals (1)-(2)".
3)
The air cooling unit functions under the "ACCELERATION" or "NORMAL
OPERATION" state.
4)
The air cooling unit stops when the STP pump stops.
*1
JIS
STP-L301/L451 Series Instruction Manual
5-3
, depth 16 mm).
*1