Table 11.4 Troubleshooting While The Stp Pump Is Rotating - Edwards STP-L301 Series Instruction Manual

Turbomolecular pumps
Hide thumbs Also See for STP-L301 Series:
Table of Contents

Advertisement

STP-L301/L451 Series Instruction Manual
11.2.5
Other Abnormalities

Table 11.4 Troubleshooting while the STP Pump is Rotating

Symptom
The STP pump stops
Insufficient ultimate
pressure
Probable causes
Failure of the backing
pump
Leakage of the piping
system
Residual molecules
11-8
Countermeasures
Check the past errors
referring to Section
11.2.2, "Displaying Past
Abnormalities/Errors and
Accumulated Running
Time".
Check the backing pump
for its capacity and start
state (use a pump with
the recommended
capacity or more).
Check the piping system
for leakage.
If the main composition is
H
or H
O, perform
2
2
baking; if it is other than
H
or H
O, clean the
2
2
inside of the vacuum
equipment.
(If gases remain inside
the STP pump, contact
Service office when it is
needed to be cleaned.)
Refer to
11.2.2
4.3.4
13.1
4.3.4
5.1
11.3.5