Attaching the STP Control Unit Front Panel to a Rack

The dimensions of the STP control unit front panel conform to EIA Standard. Therefore, this
panel can be attached to any type of commercially-available racks.
Attach the front panel unit to the rack according to the following steps:
Attach the front panel to a rack using the screw holes for the front panel.
Also, support the STP control unit from the bottom using a support angle or a
similar tool.
To protect the STP control unit during transport, remove the rubber foot from the
bottom and attach the STP control unit to the rack using the screw holes for the
rubber foot.
◇ The STP control unit cannot be supported with only the screws on the front panel
(the STP control unit is a heavy product).
Always support it from the bottom.
◇ When installing the STP control unit, DO NOT block the ventilation port of the STP
control unit by the support angle or other things. It causes the control unit to
overheat.
◇ For the dimensions of the front panel and positions of screw holes for the rubber
foot, see Section 13, "External Appearance of the STP Control Unit".