Table 10.3 Safety Functions - Edwards STP-L301 Series Instruction Manual

Turbomolecular pumps
Hide thumbs Also See for STP-L301 Series:
Table of Contents

Advertisement

STP-L301/L451 Series Instruction Manual
Operation state
Abnormality
of safety
/Error
functions
Power failure Power failure
time: 2 seconds
or longer.
Abnormal
Rotor descends
magnetic
onto the touch
bearing
down bearing.
Excessive
Rotor vibration:
vibration
100 μm 0-P or
more.
Overload of
Excessive
motor driver
current flows in
the motor.

Table 10.3 Safety Functions

LED
LCD
indication
"FAILURE"
"POWER
illuminates.
FAILURE"
"POWER"
extinguishes.
"FAILURE"
"DISTURBANCE"
"FAILURE"
"DISTURBANCE"
"FAILURE"
"OVERLOAD"
REMOTE
State of the STP pump
output signal
"ALARM"
The magnetic bearing
continues to operate using
"POWER ON"
regenerative energy.
output cut-off
The STP pump decelerates
and stops.
The rotor lands on the
touch down bearing at
approximately. 17,000
rpm.
"ALARM"
The rotor stops on the
touch down bearing.
"ALARM"
The STP pump decelerates
and stops.
"ALARM"
The STP pump decelerates
and stops.
10-4
Probable causes of an
abnormality/error
operation
Power failure.
Impossible
Power cable breakage.
Power cable disconnection.
STP connection cable
Possible
breakage.
STP connection cable
disconnection.
Abnormal control circuit.
External vibration and
Possible
impact.
Intrusion of atmospheric
air.
Intrusion of foreign
particles.
Leakage from the vacuum
Possible
equipment or piping.
Insufficient pressure at
inlet and outlet port sides.
Failure of starting of the
backing pump.
Reset by
RESET