Dp Clean High Flow Purge/Solvent Flush (Option) - Edwards EXS Series Instruction Manual

Dry screw vacuum pump
Table of Contents

Advertisement

EXS450

3.7 DP clean high flow purge/solvent flush (option)

Table 13 High flow purge/solvent flush data
Feature
Pneumatic valve gas sup-
ply
Pneumatic valve inlet con-
nection
High flow purge gas
High flow purge air filter
DP clean solvent flush flu‐
id inlet connection
DP clean solvent suction
pipe connection
Note:
The DP clean solvent suction pipe is supplied with the pump but not fitted. Refer to the
instruction given with the kit to activate the function.
A41870880_A
A41870880_A - Technical data
Pump
Ambient (°C)
40
30
20
40
15 SLM
1.8 bar dP
12 SLM
1.5 bar dP
10 SLM
1 bar dP
Specification
Nitrogen or clean dry air
3/8 inch compression fitting
Air, nitrogen or other inert gas
which is compatible with the
process
1/2 inch NPT male
(available as a spare, refer to
Table:
Spares)
3/8 inch BSP male
3/8 inch BSP female
Page 23
Water temperature (°C)
30
12 SLM
1.5 bar dP
12 SLM
1.5 bar dP
10 SLM
1 bar dP
Rating
2.5 - 6.9 barg
(36 - 100 psig)
Typically 185 slm at at-
mospheric pressure at
the purge inlet connec-
tion
07/2021 - ©Edwards Limited
20
10 SLM
1 bar dP
10 SLM
1 bar dP
10 SLM
1 bar dP

Advertisement