8. Fault finding

Table 21 Fault finding
Condition
Pump does not start up
Pump gets too hot
Motor power consumption is too high
Pump is too loud
Pump is losing lubricant
Oil gets too dark
Lubricant in the pump chamber
Pump does not attain its pumping speed
Fault
Pump does not start up
Cause
Motor incorrectly connected.
Remedy
Connect motor correctly.
Cause
Overtemperature switch or motor stator defective.
Remedy
Contact us.
Cause
Motor rotor defective.
Remedy
Contact us.
Cause
Pump has seized
‐ defective rotors, bearings or toothed gears
‐ pump restarted when it is still hot.
Remedy
Wait for 60 minutes and restart the pump again. If the pump does not start, contact
us.
Fault
Pump gets too hot
Cause
Cooling water supply is not sufficient.
Remedy
Make sure that the cooling water supply is sufficient.
Cause
Cooling water lines are clogged.
Remedy
Decalcify cooling water lines.
Cause
Ambient temperature is too high or cooling air flow is obstructed.
Remedy
Install the pump at a suitable place or ensure a sufficient flow of cooling air.
Cause
Pump is operating in the wrong pressure range.
Remedy
Check the pressure levels within the system.
Cause
Gas temperature is too high.
Remedy
Check system.