GE Brivo OEC 715 Service Manual page 82

Mobile c-arm x-ray product
Hide thumbs Also See for Brivo OEC 715:
Table of Contents

Advertisement

Brivo OEC 715/785/865/865 Mobile C-Arm X-Ray Product Service Manual
HAND_HLF_SECURITY Hand security line
termination
HAND_HLF
High Level Fluoro
hand switch
termination
HAND
High Level Fluoro
_HLU_SECURITY
hand security line
termination
HAND_COM
Common
termination
Foot Switch
Signals from foot switch to MCB are listed as follows.
Name
FOOT_FLU
Foot switch
termination
FOOT_FLU_SECURIT
Foot security line
Y
termination
FOOT_ HLF
High Level Fluoro
foot switch
termination
FOOT_
High Level Fluoro
HLF_SECURITY
foot security line
termination
FOOT_BUTTON1
Button 1
FOOT_BUTTON2
Button 2
FOOT_COM
Common
termination
Emergency Switch
Signals from emergency switch to MCB are listed as follows.
Name
EMERGENCY_SWT1 One termination of
Emergency
2-28
Passive
Passive
Passive
Passive
Description
Attribution
Passive
Passive
Passive
Passive
Passive
Passive
Passive
Description
Attribution
Passive
Open/Close
J1-C27
Open/Close
J1-B7
Open/Close
J1-A22
-
J1-C15
Scale
Pin
Open/Close
J1-A27
FOOT_FLU & FOOT_COM is
connected means FOOT_FLU
button has been pressed
Open/Close
J1-C28
-
Open/Close
J1-A14
FOOT_HLF & FOOT_COM is
connected means FOOT_HLF
button has been pressed
Open/Close
J1-C31
-
Open/Close
J1-A20
FOOT_BUTTON1 & FOOT_COM is
connected means FOOT BUTTON 1
has been pressed
Open/Close
J1-A19
FOOT_BUTTON2 & FOOT_COM is
connected means FOOT BUTTON 2
has been pressed
-
J1-B32
-
Scale
Pin
Open/Close
J1-C11
EMERGENCY_SWT1 &
REF_EMERGENCY_SWT1 is not
button has pressed
-
HAND_HLF & HAND_COM is
connected means HAND_HLF
button has been pressed
-
-
Comment
Comment

Advertisement

Table of Contents
loading

This manual is also suitable for:

Brivo oec 785Brivo oec 865