GE Brivo OEC 715 Service Manual page 114

Mobile c-arm x-ray product
Hide thumbs Also See for Brivo OEC 715:
Table of Contents

Advertisement

Brivo OEC 715/785/865/865 Mobile C-Arm X-Ray Product Service Manual
Signals from sealing board to MCB:
Name
OIL_TEMP
OIL_TEMP_GND
THERMAL_SWITCH_A
THERMAL_SWITCH_B
MA_CAT
KV_GND
Technical Specifications:
Nominal high voltage
Nominal input power
Maximum anode heat content
Target material
Target angle
Small focal spot (IEC336) (Normal Fluoro, High
Level Fluoro, Digital Spot)
Large focal spot (IEC336) (Radiography)
Filtration equiv.
Anode
Typical filament specification
Maximum filament current
Weight
Nominal heat capacity
2-60
Description
Output voltage of
temperature sensor
Reference of OIL_TEMP
One termination of
temperature switch
The other termination of
temperature switch
The cathode of mA sample
The anode of mA sample
Attribution
Scale
Comment
Analog
0-5V
-
Ground
0V
-
Passive
-
THERMAL_SWITCH_A &
THERMAL_SWITCH_B
connected means that
temperature is lower than
threshold.
THERMAL_SWITCH_A &
THERMAL_SWITCH_B
disconnected means that
temperature is higher than
threshold.
Passive
-
Analog
-
-
Analog
-
-
110 kV
2.2 kW
53 kJ
Tungsten
10°
0.6×1.4 (IEC336)
1.4 (IEC336)
0.55 mm Al
Stationary
4.2 V, 3.5A
4.6A
0.65 kg
76kHU