1. Lift up the side of the Workstation carefully and keep the
balance of the Workstation.
2. Remove the screws of the side cover and remove the side
cover.
3. Loose the nut and remove the Non-central control wheel.
4. In position 2, move the Workstation and verify that the
Workstation can be moved smoothly.
7.3.7. Monoblock With FDB Kit
During the Monoblock replacement procedure, at least 2 powerful men are required to
CAUTION
avoid C-arm unbalanced condition and injury.
After Monoblock removal, the system will be in
unbalance status. The back of C-arm will lower
than normal. Take care anything would be
pressured by C-arm. Find wood block to support the
mainframe is appreciated.
Paste the Rating Plate and Filtration Procedure
Paste the Rating plate and Filtration label in the Brivo OEC 715/785/865 label kit on the Monoblock as shown
below.
For China, paste the Chinese rating plate in the label kit.
For other countries, paste the multi-language rating plate as per local language
requirement, English rating plate is pasted as default.
Replacement
7-29