1.2.8.

Motorized Mechanical Movement

Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product has motorized mechanical assemblies. Please
follow these guidelines:
Always observe mechanical assemblies when operating the motor to avoid pinching or collision with a
person or object.
Use care when working around equipment to avoid unintentional motor actuation. Do not carelessly
place objects on the equipment or bump or lean against the equipment.
Observe and prevent articles of clothing from getting caught in moving parts.
1.2.9.

Work on Parts with Lead Materail

Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product has parts with touchable lead material. They are
Monoblock, Image Intensifier, Image Intensifier cover, lead ring of collimator and collimator cover.
When working on those parts, Do Not directly touch the lead material.
CAUTION
1.2.10.

Safety Interlock

Never bypass, jumper, or otherwise disable any equipment safety feature.
WARNING