GE Brivo OEC 715 Service Manual page 258

Mobile c-arm x-ray product
Hide thumbs Also See for Brivo OEC 715:
Table of Contents

Advertisement

1. Place the dosimeter probe on the I.I. grid center, and secure it with transparent tape.
2. Use 2mm copper to cover the beam path on the X-ray tube, and secure it with tape.
3. Position the C-arm to lateral position.
4. Make an exposure in Auto Fluoroscopy, and wait until the technique stabilizes. Check if the dose
rate is within:
Brivo OEC 715/Brivo OEC 785(I.I PN: 5075001): 4.1±0.3mR/min (35.92±2.63uGy/min or
2155±157uGy/hr)
Brivo OEC 715/785/865 (I.I PN: 5085917): 3.1±0.3mR/min (27.16±2.63uGy/min or
1629±157uGy/hr)
Brivo OEC 865(I.I PN: 5075008): 3.1±0.3mR/min (27.16±2.63uGy/min or 1629±157uGy/hr)
5. If failed, open the Camera head cover, and carefully adjust the camera iris. Loosen Camera Iris
Securing Screw, turn Camera Iris to smaller index to decrease the stable dose rate, and to bigger
index to increase stable dose rate.
Camera Iris
Adjustment Ring
Camera of Brivo OEC 715/785 5075006 Component GigE Camera AM1565
Calibration
Camera Iris
Secure Screw
4-47