Guidance and manufacturer's declaration - Electromagnetic Immunity
Brivo OEC 715/785/865
The
environment specified below. The purchaser or operator of the
ray Product should assure that it is used in an electromagnetic environment as described below:
Immunity Test
IEC 60601
Test Level
Electrostatic
±6 kV contact
discharge (ESD)
±8 kV air
IEC 61000-4-2
(Ed3.0)
±8 kV contact
±15 kV air
(Ed4.0)
±2 kV for power
supply lines
Electrical fast
transient Pulse
±1 kV for
input/output lines
IEC 61000-4-4
Surge
±1 kV Line(s) to
Line(s)
IEC 61000-4-5
±2 kV Line(s) to earth
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
0% UT
IEC 61000-4-11
(250/300 cycle)
Power frequency
3 A/m(Ed3.0)
magnetic field
(50/60Hz)
30 A/m(Ed4.0)
IEC 61000-4-8
(50Hz or 60Hz)
NOTE:
is the a.c. mains voltage prior to application of the test level.
U
T
Mobile C-arm X-ray Product is intended to use in the electromagnetic
Compliance Level
±6 kV contact
±8 kV air
(Ed3.0)
±8 kV contact
±15 kV air
(Ed4.0)
±2 kV for power supply
lines
±1 kV for input/output
lines
±1kV Line(s) to Line(s)
±2 kV Line(s) to earth
0% UT
3 A/m(Ed3.0)
30 A/m(Ed4.0)
Introduction and Safety
Brivo OEC 715/785/865
Electromagnetic
Environment Guidance
Floors are wood, concrete, or
ceramic tile, or floors are
covered with synthetic material
and the relative humidity is at
least 30 %.
Mains power quality is that of a
typical commercial and/or
hospital environment
Mains power quality is that of a
typical
commercial
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment. If the
operator of the Brivo OEC
715/785/865 Mobile C-arm X-
ray Product requires continued
operation during power mains
interruptions, it is
recommended that the Brivo
OEC 715/785/865 Mobile C-arm
X-ray Product be powered from
an uninterruptible power
supply or a battery.
Power frequency magnetic
fields are at levels
characteristic of a typical
location in a typical commercial
and/or hospital environment.
Mobile C-arm X-
and/or