If no keyboard/mouse input in 5 minutes, system will be in screen saver mode, monitor
will be turned off. Move mouse, system will be back to normal mode, monitor will be
turned on.
11. When the installation stage 2 is completed, Brivo OEC Product application install complete screen
will be displayed, Click "OK". Take out the application installation CD and shutdown system.
12. Turn on system and login APP interface to do workstation operation check as defined in section3.6.1.
13. If in step 6 select "Not retain the user data", need login Admin interface by service authority to
perform Shading Correction Calibration refer to Chapter 4 and reconfigure/set system refer to
Chapter 5.