GE Brivo OEC 715 Service Manual page 392

Mobile c-arm x-ray product
Hide thumbs Also See for Brivo OEC 715:
Table of Contents

Advertisement

Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
Exception
Error
ID
Warning
0402505
Warning
0402507
Warning
0402509
Warning
0402510
Warning
0402511
Warning
0402512
Warning
0402513
Warning
0402515
Warning
0402517
Warning
0402518
Error
6-58
Exception
Description
(Displayed in User Interface)
(In system Log)
Update exam
Update exam information
failed
failure. Please try again. If this
issue persists, Please call for
service.
Update image
Update image information
failed
failure. Please try again. If this
issue persists, Please call for
service.
DB undefined
Undefined exception
exception is
encountered. Please try again.
caught
If this issue persists, Please call
for service.
Search row in
Query information from DB
DB failed
failure. Please try again. If this
issue persists, Please call for
service.
Delete row
Delete information from DB
from DB failed
failure. Please try again. If this
issue persists, Please call for
service.
Update row in
Update information failure.
DB failed
Please try again. If this issue
persists, Please call for service.
Search exam
Query exam information from
row in DB failed
DB failure. Please try again. If
this issue persists, Please call
for service.
Search image
Query image information from
row in DB failed
DB failure. Please try again. If
this issue persists, Please call
for service.
Insert row
Insert information failure.
failed
Please try again. If this issue
persists, Please call for service.
DatabaseMana
Workstation communication
gement module
exception. Please restart the
register servant
system. If this issue persists,
exception
please call for service.
Message
Diagnosis Reference
For example: updateExam()
1. getExam(examId);--> not
found;