GE Brivo OEC 715 Service Manual page 448

Mobile c-arm x-ray product
Hide thumbs Also See for Brivo OEC 715:
Table of Contents

Advertisement

2. Disconnect the cable connected with the
membrane keyboard. Remove the front upper
cover.
3. Remove the screws that secure the membrane
keyboard.
4. Reinstall a new Workstation control panel kit.
5. Reassemble the membrane keyboard in the new
upper cover.
6. Remove the screws that secure the rear upper
cover. Remove the rear upper cover.
7. Replace the new upper cover kits and install them
in reverse sequence.
The torque of the screws that secure the top covers is 2.5Nm.
Replacement
7-41