Auto Technique Tracking - GE Brivo OEC 715 Service Manual

Mobile c-arm x-ray product
Hide thumbs Also See for Brivo OEC 715:
Table of Contents

Advertisement

2. Place X-ray Keyswitch in XRAYOFF position (
are generated. Press vertical lift button and verify the vertical lift column is operational.
3. Place X-ray Keyswitch in ON position (
to verify X-ray and vertical lift column are enable.
Emergency Switch
Perform the following test to verify the emergency switch.
1. Press the emergency switch during raising C-Arm.
Emergency Switch
2. Verify that vertical column operation and exposure are disabled.
3. Release the emergency switch. Verify that the vertical column operation and exposure are
available.
3.6.3.

Auto Technique Tracking

Select Auto Fluoroscopy mode (
filter on the Monoblock head. Verify that the kVp changes accurately as copper filters are inserted into
the field in the following ranges:
Model
Brivo OEC 715/785
(I.I PN: 5075001):
Brivo OEC 715/785/865
(I.I PN: 5085917):
Brivo OEC 865 (I.I PN: 5075008):
If the kVp is not in the range, refer to 4.6.7 for calibration.
). Press Handswitch and verify that no X-rays
). Press Handswitch, footswitch and vertical lift button
), and the NORM field size to take an exposure. Put the copper
Copper Filter Thickness
kVp range
kVp range
kVp range
1mm
2mm
64±4kVp
74±4kVp
62±4kVp
72±4kVp
62±4kVp
72±4kVp
Installation
3mm
82±4kVp
80±4kVp
80±4kVp
3-45