28A (200V)
•
25A (220V/230V/240V)
Rated Continuous Line Current:
•
20A (100V/110V/120V)
•
12A (200V)
•
10A (220V/230V/240V)
Maximum line impedance: 0.3
Max Line Regulation (Percent): 15% at maximum radiographic exposure.
Line regulation is based on measurements made at the input (primary winding) of the isolation
transformer under standby conditions and at maximum radiographic exposure by the following
equation:
Line regulation = 100 (Vn - V1)/V1
Where Vn = no load voltage (standby), and V1 = loaded voltage (maximum radiographic exposure).
16A power supply is required in China. And 20 A is required in Japan.
9.2.4.
Circuit Breaker
Breaker for 110V:
•
Rated Voltage
•
Frequency
•
Rated Current
Breaker for 220V:
•
Rated Voltage
•
Frequency
•
Rated Current
9.2.5.
Earth leakage current
Normal condition
Single fault condition
Ω
Ω
Ω
|0.6
|0.6
AC 250 V
50 Hz /60 Hz
20A (AC)
AC 250 V
50 Hz /60 Hz
12A (AC)
0.5mA
0.5mA