Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
Exception
Error
ID
Warning
0306505
Warning
0306506
Warning
0306507
Warning
0306508
Warning
0306509
Warning
0306510
Warning
0306511
Warning
0306513
Warning
6-24
Exception
Description
(Displayed in User Interface)
(In system Log)
Load abandon
Abandon Reason loading
reason list
failure. Abandon Exam
failed
function will be impacted.
Please restart the system. If
this issue persists, please call
for service.
Load DICOM
Connectivity configuration
printer list
loading failure. DICOM printer
failed
list missed. Please restart the
system. If this issue persists,
please call for service.
Load setup
Setup configuration loading
configuration
failure. Some setting may be
file failed
missed. Please restart the
system. If this issue persists,
please call for service.
Load exam list
Setup configuration loading
option failed
failure. Exam list display setting
missed. Please restart the
system. If this issue persists,
please call for service.
Load default
Connectivity configuration
MPPS server
loading failure. Default MPPS
failed
server setting missed. Please
restart the system. If this issue
persists, please call for service.
Load DICOM
Connectivity configuration
storage server
loading failure. Storage server
failed
setting missed. Please restart
the system. If this issue
persists, please call for service.
Read local AE
Connectivity configuration
title failed
loading failure. Local AE title
missed. Please restart the
system. If this issue persists,
please call for service.
Read MWL
Connectivity configuration
server list failed
loading failure. Worklist server
list missed. Please restart the
system. If this issue persists,
please call for service.
Message
Diagnosis Reference
1. Check mainboard of
workstation
2. Check display card of
workstation
3. Reinstall workstation
software.
1. Restart the system.
2. If not work, re-install
Workstation software.
1. Restart the system.
2. If not work, re-install
Workstation software.
1. Restart the system.
2. If not work, re-install
Workstation software.
1. Restart the system.
2. If not work, re-install
Workstation software.
1. Restart the system.
2. If not work, re-install
Workstation software.
1. Restart the system.
2. If not work, re-install
Workstation software.