9.5. Radiation Safety Specification

Operators are strongly urged to be familiar with the current recommendations from the International
Commission on Radiological Protection. (ICPR)
Take radiation protection of the patient and operator into consideration seriously.
Scatter data in this manual is only for reference.

9.5.1. Scatter Data

Filter in the X-Ray tube assembly
Additional filter
Total filter (without removable filter)
Leakage radiation
SID (focal spot to image receptor distance)
Patients must be kept as far as possible from the focus, with minimum distance of 20cm.
Make sure not to operate when a plastic cover in x-ray assembly is disassembled, which
WARNING
properly make the distance meet the above requirements.
The operating table matched this equipment should be made of small attenuation
equivalent material, otherwise, X-ray dose may be increased.
9-14
3.35mm Al
0mmAl
3.35mmAl
Dose area product rate is less than
1.0mGy/h, averaged over any area of
100cm
of which no principal linear
2
P
P
dimension exceeds 20cm at 1m from the
focal spot under conditions of loading.
(110kVp, 2mA)
100cm