3.5.5.

System Configuration in Admin Interface

Setup in Admin Interface by Service Authority
1. Hold down "Ctrl +Shift+Alt" and"→ " keys at same time to invoke Admin Interface. Select Service
as User, and then input password to Login Service Tool Interface.
2. Input the system ID of the system. Refer to section 5.2.1.
3. Activate Product Mode configuration and software option feature by inputting e-License key,
refer to section 5.2.2.
Before activate Product Mode configuration by inputting e-License key, an error message
will be prompted as below:
4. Check Fluoro Alarm Control through exposure management screen. Refer to section 5.2.4.
5. Check the exposure option and dose group by country through Exposure Management screen.
Refer to section 5.2.4.
The configuration must be complete in case the local regulation is not complied.
CAUTION
6. Click "Shutdown" button to shut down the system.
Setup in User Setup interface
3-22