Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
1.
Remove the image intensifier refer to the bundle cable replacement instruction.
2.
Loosen four screws and remove the I.I plate, and then pull out from the C-arm.
3.
Loosen the brake by enclasping the C-arm with at least 2 people's effort and pull out the truck along
the orbital direction, then place the C-arm with generator on the soft underlay.
4.
Replace the Rolling Wheels with Retainer ring clamp.
5. Reassemble and recover the system per the reverse order.