4.2. kV Measurement and Adjustment

4.2.1. kV Measurement and Accuracy Check

The X-ray generator is composed of high voltage transformer, high voltage rectifier and X-ray tube in one
pressurized oil tank. kV value can be measured by kV meter. Position the kV meter in the center of the X-
ray beam 30 cm above the focal spot.
KV Accurancy requirement
Take exposure at kV values below in manual normal Fluoro mode. The error should be within ±7%
ranges. Otherwise, perform the kV adjustment procedure in section 4.2.2 until all the kV values meet the
requirement below.
Exposure
kVp
mode
value
Low Dose
99 kVp
Normal
66kVp
Fluoro
110 kVp
High Level
55 KVp
Fluoro
110 kVp
Digital Spot
55 KVp
80 kVp
110 kVp
Radiography 55 kVp
80KVp
110 kVp
KV Accurancy requirement in Australia
Take exposure at kVp values below in normal fluoro mode. The error should be within ±5% ranges.
Otherwise, perform the kVp adjustment procedure in Section 4.2.2 until all the kVp values meet the
requirements below.
Exposure
kVp
mode
value
Low Dose
99 kVp
Normal
66kVp
Fluoro
110 kVp
mA or mAs Set Value
2mA
4mA
1mA
4mA
4mA
4mA
4mA
4mA
40mAs (For 100/110V/120V input)
80mAs (For
200V/220V/230V/240V input)
40mAs
20mAs
mA or mAs Set Value
2mA
4mA
1mA
Calculation
Minimum
Error
(kVp)
7%
92.07
7%
61.38
7%
102.3
7%
51.15
7%
102.3
7%
51.15
7%
74.4
7%
102.3
7%
51.15
7%
74.4
7%
102.3
Calculation
Minimum
Error
(kVp)
5%
94.05
5%
62.7
5%
104.5
Calibration
Maximum
(kVp)
105.93
70.6
117.7
58.85
117.7
57.75
85.6
117.7
58.85
85.6
117.7
Maximum
(kVp)
103.95
69.3
115.5
4-3