Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
DAP Unit Config:
Field engineer can configure software DAP unit on this interface.
HLF level exposure setting:
Disable after High Level Exposure: Enable or disable "The system automatically Disable After High Level
Exposure" function. Some countries or regions, such as certain provincial locations in Australia, may
require enabling it.
Disable after No Exposure for 5 Minutes: Enable or disable "Disable after No Exposure for 5 Minutes"
function. Some countries or regions, such as certain provincial locations in Australia, may require
enabling it.
HLF Accessible Only in Auto ABS: Enable or disable "HLF Accessible Only in Auto ABS" function. Some
countries or regions, such as certain provincial locations in Australia, may require enabling it.
Automatic Termination at 20" in One Shot: Enable or disable "Automatic Termination at 20" in One Shot"
function. Some countries or regions, such as Australia or Switzerland, require enabling it.
Refer to the following table "Exposure Management Matrix" to set HLF scene setting.
Countries /
Dose Group Option
Regions
Albania
Normal 88 and HLF 100
Algeria
Normal 88 and HLF 100
Angola
Normal 88 and HLF 100
Argentina
Normal 88 and HLF 100
Australia
Normal 50 and HLF 100
Austria
Normal 88 and HLF 100
Bahrain
Normal 88 and HLF 100
Bangladesh Normal 88 and HLF 100
Belarus
Normal 88 and HLF 100
Belgium
Normal 88 and HLF 100
Bolivia
Normal 88 and HLF 100
Bosnia
Normal 88 and HLF 100
Brazil
Normal 50 and HLF 100
Brunei
Normal 88 and HLF 100
5-8
Exposure Management Matrix
HLF
Automatic
Accessible
Termination
Only in
at 20" in
Auto ABS
One Shot
Disable
Enable
Disable
Enable
Disable
Enable
Disable
Enable
Enable
Enable
Disable
Enable
Disable
Enable
Disable
Enable
Disable
Enable
Disable
Enable
Disable
Enable
Disable
Enable
Disable
Enable
Disable
Enable
Disable after
Disable after
No Exposure
High Level
for 5
Exposure
Minutes
Disable
Disable
Disable
Disable
Disable
Disable
Disable
Disable
Enable
Enable
Disable
Disable
Disable
Disable
Disable
Disable
Disable
Disable
Disable
Disable
Disable
Disable
Disable
Disable
Disable
Disable
Disable
Disable
Auto
Termination
at 9'30" in one
shot
Enable
Enable
Enable
Enable
Enable
Enable
Enable
Enable
Enable
Enable
Enable
Enable
Enable
Enable

Advertisement

Table of Contents
loading

This manual is also suitable for:

Brivo oec 785Brivo oec 865

Table of Contents