Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6. Replace the new Image intensifier on C-Arm.
Be carefully when disassemble the Image
CAUTION
intensifier and make cables pass through
the Image intensifier housing to avoid
Image Intensifier drop.
Be careful not to touch the camera lens or
CAUTION
image intensifier output window and keep
both of them clear without any impurity.
7. Reassemble Image intensifier power supply, CCD camera, counter weight and connect all cables.
Ensure the cable tightly connect to the Image intensifier Power Supply. False
connection will lead to no image or image vibration in Fluoro.
8. Refer to 7.3.2 calibration matrix and chapter 4 to do calibration and checks.
9. After calibration and check, shut down system and reinstall covers removed in above steps.

7.3.15. Image Intensifier Power supply

There are 2 FRUs for Image intensifier power supply (P/N: 5075009 for 5075929/5075930 and
5085918 for 5075967).
1.
Remove Image intensifier cover, and then disconnect all cables connected to the power supply
and CCD Camera. Remove secure screws of power supply on Image intensifier for
5075929/5075930. Refer to 7.3.14 for the Image Intensifier removal procedure for 5075967.