Replacement
Replacement procedure:
1. Move C-arm to a clear area.
2. Remove mainframe front and back covers and then open mainframe package.
3. Cut all the bundle cable related cable ties in Mainframe base
4. Disconnect all related cables.
5. Remove 3 securing screws to fix interconnect cable bracket.
6. Remove two cable sleeves.
7. Refer to section 7.3 remove Monoblock, collimator and Image Intensifier.
8. Remove the Monoblock support and pull out all cables via C-arm carefully.