a. NORM field size: Set Field size to NORM, adjust trimmer A31.
b. MAG1 filed size: Set Field size to MAG1, adjust trimmer A32.
c. MAG2 filed size: Set Field size to MAG2, adjust trimmer A33.
3. Take an exposure, and check the Field size, repeat as many times as necessary as field size is in
specification.
Cleanup
1. After passing, remove the BAFT/UBAT from the I.I. Grid.
2. Make sure to re-assembly of all components as the following
a. Power off the system.
b. Disconnect Camera Cables.
c. Reinstall the I.I. Power Supply and connect cables between power supply and I.I.
4-44