Advertisement

5. mA accuracy in Radiography Mode:
Measure the voltage of TP7 (MA-SMP) on MCB board with oscilloscope:
mA measured=Sample Voltage (TP7) (unitV)*1000/200ohm
mA real=mA accuracy measure
mA accuracy=[(mA set – mA actual )/mA set]*100%
Set the exposure mode and parameter to check to the voltage (TP7) values. The voltage of TP7 should be
in the range of the form below. Otherwise, mA calibration procedures should be performed until all the
voltage values meet the requirement in the list.
For power input: 100V/110V/120V (Setting in Exposure Management)
Exposure mode KV and mAs value setting
Radiography
40KVp@40mAs
110KVp@40mAs
For power input: 200V/220V/230V/240V (Setting in Exposure Management)
Exposure mode KV and mAs value setting
Radiography
40KVp@80mAs
110KVp@40mAs
mA value
Voltage (TP7)
minimum (V)
10mA
-2.20
10mA
-2.20
mA value
Voltage (TP7)
minimum (V)
20mA
-4.40
20mA
-4.40
Calibration
Voltage (TP7)
maximum (V)
-1.80
-1.80
Voltage (TP7)
maximum (V)
-3.60
-3.60
4-11