Exception
Error
ID
Warning
0304511
Warning
0304512
Warning
0304513
Warning
0304514
Warning
0304516
Warning
0304517
Warning
0304518
Warning
0304519
Warning
Exception
Description
(Displayed in User Interface)
(In system Log)
Set current
Set current exam failure.
exam to
Please restart the system. If
Imaging Engine
this issue persists, please call
failed
for service.
Delete image
Delete exam exception. Please
by exam tco
try again. If this issue persists,
failed
please call for service.
Get text
Get text resource failure.
resource failed
Application display may be
abnormal. Please restart the
system. If this issue persists,
please call for service.
Load
Connectivity configuration
connectivity
loading failure. Some
configuration
configuration may be missed.
file fail
Please restart the system. If
this issue persists, please call
for service.
Load default
Connectivity configuration file
storage server
loading exception. Default
failed
storage server setting is
missed. Please restart the
system. If this issue persists,
please call for service.
Init exam list
Setup configuration loading
option failed
failure. Exam list display setting
missed. Please restart the
system. If this issue persists,
please call for service.
Init height &
Setup configuration loading
weight unit
failure. Weight/Height units
failed
missed. Please restart the
system. If this issue persists,
please call for service.
Init default
Connectivity configuration
MPPS server
loading failure. Default MPPS
failed
server setting missed. Please
restart the system. If this issue
persists, please call for service.
Message
Troubleshooting
Diagnosis Reference
1. Restart the system.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.
1. Restart the system.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.
1. Restart the system.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.
1. Restart the system.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.
1. Restart the system.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.
1. Restart the system.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.
1. Restart the system.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.