Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
Exception
Error
ID
Warning
0309508
Warning
0309509
Warning
0309510
Warning
0309511
Warning
0309512
Warning
0309513
Error
0309514
Error
0309515
Error
0309516
Error
0309517
Error
6-40
Exception
Description
(Displayed in User Interface)
(In system Log)
Start setup
Open Setup exception. Please
exception
restart the system. If this issue
persists, please call for service.
Stop setup
Close Setup exception. Please
exception
restart the system. If this issue
persists, please call for service.
Read
Connectivity configuration
configuration
loading failure. Please restart
file failed
the system. If this issue
persists, please call for service.
Init tab page
Runtime exception. Please
controller failed
restart the system. If this issue
persists, please call for service.
Read serial
Workstation hardware version
number from
access exception. Hardware
bios failed
information can't be displayed.
Please restart the system. If
this issue persists, please call
for service.
Register CORBA
Workstation communication
service failed
exception. Please restart the
system. If this issue persists,
please call for service.
Unregister
Workstation communication
CORBA service
exception. Please restart the
failed
system. If this issue persists,
please call for service.
CORBA sync
Workstation communication
notify failed
exception. Please restart the
system. If this issue persists,
please call for service.
Subscribe sync
Workstation communication
event failed
exception. Please restart the
system. If this issue persists,
please call for service.
Unsubscribe
Workstation communication
sync event
exception. Please restart the
failed
system. If this issue persists,
please call for service.
Message
Diagnosis Reference
system unknown error
system unknown error
check if configuration file
crashed
system unknown error
1. Restart the system.
2. If not work, re-install
Workstation software.
make sure if omniNames
service failed, any module
initialization failed
make sure if omniNames
service failed, check
omniOrb log file
make sure if omniNames
service failed,
CorbaEventChannel
initialization failed
make sure if omniNames
service failed,
CorbaEventChannel
initialization failed
make sure if omniNames
service failed,
CorbaEventChannel
initialization failed