Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6.1. Overview
This chapter describes messages that appear on the right workstation monitor during system operation.
Messages may indicate any of the following:
•
Error Messages
•
Warning Messages
•
Information
When the error message appears, error code will also be displayed on the kV, mA display field on control
panel, if the Mainframe is connected with Workstation.
6.2. Error Recovery Steps
Perform the following error recovery procedures if you encounter problems during start-up or operation:
1. Some messages can be cleared by pressing OK button or taking exposure, but will periodically re-
display while the condition exists.
2. If an error message occurs, shutdown the system. Wait one minute, and then press the power on key.
If this fails to restore normal operation then proceed with step 3.
3. Press the power off key, disconnect the power cord, and call for service. Refer to chapter 1 for
communication center telephone numbers. Do not continue using the system.
System may fail at any time. Ignoring error and warning messages may result in
WARNING
equipment damage and personal injury.
6.3. Event Messages
6.3.1.
Structure of Message Code
Three subcodes constitute the message:
Subsystem Code
Component Code
**
Example:
01
6-2
Exception No.
**
***
02
001
Message ID
**_**_***
01 02 001
Tube temperature sensor
feedback SPI line work
abnormal
Description
-------