GE Brivo OEC 715 Service Manual page 389

Mobile c-arm x-ray product
Hide thumbs Also See for Brivo OEC 715:
Table of Contents

Advertisement

Exception
Error
ID
Warning
0405707
Warning
0405708
Error
0405709
Error
0407501
Error
0407502
Error
0407503
Warning
0407504
Error
0407505
Error
0407506
Error
0407507
Error
Exception
Description
(Displayed in User Interface)
(In system Log)
SCU module
Workstation communication
register servant
exception. Please restart the
exception
system. If this issue persists,
please call for service
SCU inits a
Semaphore exception. Please
semaphore
restart the system. If this issue
failed
persists, please call for service.
SCU module
Workstation communication
unregister
exception. Please restart the
servant
system. If this issue persists,
exception
please call for service
MM module
Workstation communication
resolve object
exception. Please restart the
exception
system. If this issue persists,
please call for service
MM module
Workstation communication
Invoke corba
exception. Please restart the
service
system. If this issue persists,
exception
please call for service
Cancel thread
Cancel execution thread
failed
exception. Please restart the
system. If this issue persists,
please call for service.
MM module
Workstation communication
register servant
exception. Please restart the
exception
system. If this issue persists,
please call for service
Init semaphore
Semaphore exception. Please
failed
restart the system. If this issue
persists, please call for service.
Set locale failed
System start failure. Set locale
failed. Please restart the
system. If this issue persists,
please call for service.
Allocate
System run out of memory.
memory failed
Please restart the system. If
this issue persists, please call
for service.
Message
Troubleshooting
Diagnosis Reference
make sure if omniNames
service failed, any module
crashed
system unknown error
Make sure if omniNames
service failed, any module
initialization failed.
Make sure if omniNames
service failed, check
omniOrb log file.
System error: can't cancel a
thread.
Make sure if omniNames
service failed, check
omniOrb log file.
Check if the semaphore is
valid, or check if the
maximum allowable value
for the semaphore will be
exceeded.
System error. Call to
setlocal() function failed.
Check memory using state.
6-55

Advertisement

Table of Contents
loading

This manual is also suitable for:

Brivo oec 785Brivo oec 865

Table of Contents