GE Brivo OEC 715 Service Manual page 384

Mobile c-arm x-ray product
Hide thumbs Also See for Brivo OEC 715:
Table of Contents

Advertisement

Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
Exception
Error
ID
Warning
0411801
Warning
0411802
Warning
0403501
Error
0403502
Error
0403503
Warning
0403504
Warning
0403505
Warning
0403506
Warning
6-50
Exception
Description
(Displayed in User Interface)
(In system Log)
System's image
The image can't be saved.
capacity reach
Image storage capacity limit of
limit
100,000 images reached.
Please delete useless or out-of-
date images to make disk
space for new images.
Exam's image
Image storage capacity limit of
capacity reach
9999 images for each exam
limit
reached. No more images can
be saved under the exam.
Please delete some useless
images if you want to save
new images of the exam.
DTR sets locale
System start failure. Set locale
failed
failed. Please restart the
system. If this issue persists,
please call for service.
DTR reports
System start failed. Please
ready failed
restart the system. If this issue
persists, please call for service.
DTR parse
Parse retrieved image failure.
retrieved image
Please try again. If this issue
failed
persists, please call for service.
DTR gets exam
Parse exam information of raw
information
images failure. Please try
failed
again. If this issue persists,
please call for service.
DTR creates
Create DICOM RDSR failure.
RDSR object
Please try again. If this issue
failed
persists, please call for service.
DTR creates
Create DICOM file exception.
DICOM file
Please try again. If this issue
setup failed
persists, please call for service.
Message
Diagnosis Reference
remove some images