KERN CKE 360-3 Operating Manual

KERN CKE 360-3 Operating Manual

Counting balances cke/cds series
Hide thumbs Also See for CKE 360-3:

Advertisement

Operating Manual
Counting balances
KERN CKE/CDS
Version 2.6
01/2015
GB
KERN & Sohn GmbH
Ziegelei 1
D-72336 Balingen
email: info@kern-sohn.com
Phone: +49-[0]7433- 9933-0
Fax: +49-[0]7433-9933-149
Internet: www.kern-sohn.com
CKE-BA-e-1526

Advertisement

Table of Contents
loading

Summary of Contents for KERN CKE 360-3

  • Page 1 KERN & Sohn GmbH Ziegelei 1 Phone: +49-[0]7433- 9933-0 D-72336 Balingen Fax: +49-[0]7433-9933-149 email: info@kern-sohn.com Internet: www.kern-sohn.com Operating Manual Counting balances KERN CKE/CDS Version 2.6 01/2015 CKE-BA-e-1526...
  • Page 2: Table Of Contents

    Version 2.6 01/2015 Operating Manual Counting balances Table of Contents MODE – MENU ....................... 3 Technical Data ....................... 4 KERN CKE ............................4 KERN CDS ............................7 Basic Information (General) .................. 9 Proper use ............................9 Improper Use ............................9 Warranty.............................. 9 Monitoring of Test Resources ......................
  • Page 3: Mode - Menu

    1 MODE – MENU How to invoke the mode menu: Turn on balance, press and hold the tare key and press the ON/OFF key. Let go of the tare key. / CDS Factory settings: 9600bd: Fast: Keyboard overview / Function Using the automatic reference optimization (OPT) the counting accuracy is automatically increased when applying up to 100 items.
  • Page 4: Technical Data

    2 Technical Data 2.1 KERN CKE KERN CKE 360-3 CKE 2000-2 CKE 3600-2 CKE 6K0.02 Readability (d) 0,001 g 0,01 g 0,01 g 0,02 g Weighing range (max) 360 g 2.000 g 3.600 g 6.000 g Taring range (subtractive) 360 g 2.000 g...
  • Page 5 KERN CKE 8K0.05 CKE 16K0.05 CKE 16K0.1 Readability (d) 0,05 g 0,05 g 0,1 g Weighing range (max) 8.000 g 16.000 g 16.000 g Taring range (subtractive) 8.000 g 16.000 g 16.000 g Reproducibility 0,05 g 0,1 g 0,1g Linearity ±0,15 g...
  • Page 6 KERN CKE 36k0.1 CKE 65k0.2 CKE 65k0.5 Readability (d) 0,1 g 0,2 g 0,5 g Weighing range (max) 36.000 g 65.000 65.000 Taring range (subtractive) 36.000 g 65.000 65.000 Reproducibility 0,2 g 0,4 g 0,5 g Linearity ±0,5 g ±1,0 g ±...
  • Page 7: Kern Cds

    2.2 KERN CDS KERN 15K0.05 16K0.1 30K0.1 30K0.1L 4K0.02 Readability (d) 0,02 g 0,05 g 0,1 g 0,1 g 0,1 g Weighing range (max) 4.000 g 15.000 g 16.000 g 30.000 g 30.000 g Taring range (subtractive) 4.000 g 15.000 g 16.000 g...
  • Page 8 KERN CDS 36K0.2L CDS 60K0.2 CDS 100K0.5 Readability (d) 0,2 g 0,2 g 0,5 g Weighing range (max) 36.000 g 60.000 g 100.000 g Taring range (subtractive) 36.000 g 60.000 g 100.000 g Reproducibility 0,2 g 0,4 g 0,5 g Linearity ±...
  • Page 9: Basic Information (General)

    The responsible user must define a suitable interval as well as type and scope of this test. Information is available on KERN’s home page (www.kern-sohn.com) with regard to the monitoring of balance test substances and the test weights required for this. Our accredited DKD calibration laboratory offers fast and inexpensive adjustment for test weights and weighing balances (reset to national normal weight).
  • Page 10: Basic Safety Precautions

    4.1 Pay attention to the instructions in the Operation Manual Carefully read this operation manual before setup and commissioning, even if you are already familiar with KERN balances. Versions in other languages are non-binding translations. The only binding version is the original document in German.
  • Page 11: Unpacking, Setup And Commissioning

    The balance must be installed in a way that the weighing plate is exactly in horizontal position. 6.3 Mains connection Power is supplied via the external mains adapter. The stated voltage value must be the same as the local voltage. Only use original KERN mains adapters. Using other makes requires consent by KERN. CKE/CDS-BA-e-1526...
  • Page 12: Battery Operation Cke

    Before connecting or disconnecting of additional devices (printer, PC) to the data interface, always disconnect the balance from the power supply. With your balance, only use accessories and peripheral devices by KERN, as they are ideally tuned to your balance.
  • Page 13: Adjustment

    6.7 Adjustment As the acceleration value due to gravity is not the same at every location on earth, each balance must be coordinated - in compliance with the underlying physical weighing principle - to the existing acceleration due to gravity at its place of location (only if the balance has not already been adjusted to the location in the factory). This adjustment process must be carried out during the initial start-up, after change in location and variation of surrounding temperature.
  • Page 14: Operation

    7 Operation 7.1 Control panel CKE/CDS ON/OFF Taring; Invoke the mode menu by pressing the ON/OFF key. Generate reference using 5 parts; In mode menu: NO function Generate reference using 10 parts Generate reference using 20 parts; In mode menu: YES function Switchover pcs ↔...
  • Page 15: Operation

    7.2 Operation 7.2.1 Adjusting 7.2.2 Speed 7.2.3 Auto Off CKE/CDS-BA-e-1526...
  • Page 16 CKE/CDS-BA-e-1526...
  • Page 17: Display Background Illumination

    7.2.4 Display background illumination Switch on the balance and the zero display, then call-up the balance menu as specified in chapter 1. Select menu item „Backlight“ using the button necessary for the respective model. Acknowledge by pressing the „YES“ button in order to switch on the background illumination permanently.
  • Page 18: Data Output Rs 232 C

    Baud rate selectable from 2400, 4800, 9600 Baud (factory setting) and 19200 Baud.  Sub-D plug 9-channel required  For operation with interface faultless operation is only ensured with the correct KERN – interface cable (max. 2m) Pin allocation of the balance output socket (front view) Sub-D jack 9-channel...
  • Page 19: Interface Rs 232C

    7.4 Interface RS 232C Data output via interface RS 232C General Information The previous condition for the data transfer between balance and a peripherical device (e.g. printer, PC ...) is that the appliances are set to the same interface parameters (e.g. baud rate, parity ...). 7.4.1 There are 4 kinds of data output via RS 232C Data output using the PRINT key The printing process can be triggered by pressing the PRINT key.
  • Page 20: Printer

    Printer The serial interface RS 232 facilitates the connection of a printer. The printout shows the weight in grams. In counting mode either the piece number or the weight details will be printed out. In percentage mode the percentage proportion or the weight details will be printed out. Printout will take place after pressing the PRINT key.
  • Page 21: Service, Maintenance, Disposal

    Spilled weighing goods must be removed immediately. 8.2 Service, maintenance The appliance may only be opened by trained service technicians who are authorized by KERN. Before opening, disconnect from power supply. 8.3 Disposal Disposal of packaging and appliance must be carried out by operator according to valid national or regional law of the location where the appliance is used.
  • Page 22: Declaration Of Conformity

    EN 50581:2012 Datum 17.07.2014 Signatur Date Signature Ort der Ausstellung 72336 Balingen Albert Sauter Place of issue KERN & Sohn GmbH Geschäftsführer Managing director KERN & Sohn GmbH, Ziegelei 1, D-72336 Balingen, Tel. +49-[0]7433/9933-0 Fax +49-[0]7433/9933-149, E-Mail: info@kern-sohn.com, Internet: www.kern-sohn.com CKE/CDS-BA-e-1526...

Table of Contents