13
Errors and faults
13.1
Error messages
Error message
L0
e
13.2
Faults
In case of a failure in the program sequence, briefly turn off the balance and disconnect
from power supply. The weighing process must then be restarted from the beginning.
Fault
Display is not lit up
The balance cannot be switched
on
The display does not react to load
change
The displayed weight is
permanently changing
The weighing result is obviously
incorrect
CH-BA-e-2440
Explication
Capacity of batteries exhausted
•
After switching on: Weighing cell defective
•
During a weighing process: Overload
•
During adjustment: Faulty adjustment (adjustment weight exceeded or
too low)
Possible cause
•
The balance is not switched on.
•
The battery was not plugged.
•
This battery is empty.
•
Battery empty or defective
•
[ON/OFF/TARE] defective
•
[ON/OFF/TARE] not correctly activated
•
Load cell defective
•
Load cell cable defective
•
Draught/air movement
•
Vibration at the hook
•
Suspended load is moving
•
The hook has contact with foreign objects
•
Electromagnetic fields / static charging (choose different
location/switch off interfering device if possible)
•
Load cell defective
•
The balance display is not at zero when balance unloaded
•
Adjustment has been carried out with a wrong or an
inaccurate adjustment weight
•
Wrong weighing unit selected
•
Electromagnetic fields / static charging (choose different
location/switch off interfering device if possible)
Errors and faults
en | 43