1) Using the mechanical measuring device, measure and record the
Mechanical Length at Zero Load "L
Note: The following steps assume that the user has performed the steps in the
previous section and calibrated the velocity. Therefore, the velocity has been
adjusted, approximate length already entered, and the MAX II is currently setup
and ready to measure in elongation mode.
2) Apply a drop of couplant to the bolt or transducer, and attach it to one end of
the bolt. Rotate the transducer clockwise and counter clockwise applying a
small amount of pressure to eliminate any excess couplant between the
transducer and bolt. Be sure to always place the transducer in the same
location. This will help to eliminate any potential measurement errors cause
by changing the sound path.
3) Record the Ultrasonic Length at Zero Load "L
4) Repeat step 2 & 3 for each sample bolt.
5) Place the first sample bolt in the loading apparatus. Apply approximately
1/3 of the maximum load, which is to be placed on the bolt under actual
Find Quality Products Online at:
L
SonicStressFactor
L
L
UltrasonicLength
U 0
L
MechanicalLength
M 0
L
UltrasonicLength
Ul
L
MechanicalLength
Ml
Calibrating the Stress Factor
GlobalTestSupply
www.
L
L
0
MECHANICAL
Ml
M
L
L
0
ULTRASONIC
Ul
U
@
ZeroLoad
@
ZeroLoad
@
Load
@
Load
" for each sample bolt.
M0
".
U0
82
.com